Burst Not Found - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Condition Messages
Err#
Bit in
Status
Register
207
2
207
2
209
3
Instrument Messages
Message

Burst Not Found

Burst not found;with
selected Time Slot
Timing Error
Error or
More Information
Warning
E
The burst signal cannot be detected because
of inappropriate parameter settings or
incorrect signal content:
— An inappropriate parameter setting
could cause the signal to be
partially, rather than fully, on the
display, Burst Search Threshold
and/or Burst Search Length may
need to be adjusted.
— An incorrect signal could have
either insufficient power, the rising
or falling edges cannot be detected,
or the burst is less than
126 microseconds.
— Carrier signal might not be actually
bursted.
W-CDMA: Either the signal being analyzed
has insufficient power, the rising or falling
edges cannot be detected, or the burst is less
than 126 microseconds.
W_CDMA: Cannot synchronize measurement
with PRACH channel for Power Control
measurement, because the signal cannot be
found. Make sure PRACH is present in the
W-CDMA uplink signal, and that the preamble
signature and scramble code are set
correctly.
GSM: Data was acquired but a GSM burst was
not found, with the timeslot mode disabled.
Bluetooth: The burst that has been found
does not correspond to the currently selected
Bluetooth packet type (the burst length may
be too short).
WLAN: The instrument cannot find a valid
WLAN burst. You may need to extend the
search length.
In the PSA, this error was reported as one of
the following error numbers: 10772, 13104,
10160, 10286, 10420, 10454, 10614, 10904,
10928, 13074, 10287
E
The selected timeslot does not contain the
expected burst.
E
99

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents