Instrument Messages
Event Messages
Err#
Message
-315
Configuration memory
lost
-320
Storage fault;
-321
Out of memory
-321
Out of memory;
Results truncated
-321
Out of memory;
Sequence too long
-321
Out of memory; Total
analysis step reaches
its limitation
-340
Calibration failed;
<failure msg><port>
-350
Queue overflow
-360
Communication error
-360
Communication error;
SNS data read
failure. Disconnect
then reconnect SNS
-360
Communication error;
SNS is not connected
-361
Parity error in
program message
-362
Framing error in
program message
-363
Input buffer overrun
-365
Time out error
Instrument Messages
Verbose/Correction Information
The nonvolatile configuration data saved by the instrument has been
lost.
A problem was found while using data storage. The error is not an
indication of physical damage or failure of any mass storage element.
An internal operation needed more memory than was available. Report
this error to the nearest Keysight Technologies sales or service office.
Reduce the acquisition count, analysis step count, or measurement
count.
The Sequence is too large to be captured. Reduce the Number of
Acquisitions or the Acquisition Duration of each Acquisition.
Reduce the analysis steps count.
The calibration for one of the I-Q ports did not succeed. The information
in the "failure msg" field can be used to troubleshoot this problem.
Contact Keysight technical support.
An error occurred that did not get put in the error queue because the
queue was full.
There was a problem with instrument remote communications. The
exact problem cannot be specifically identified.
The Keysight Smart Noise Source connected to the MXA has failed to be
read by the application. Please disconnect and reconnect the SNS. If
this continues to fail, then the SNS may have had its EEPROM corrupted
or another hardware fault exists. Check SNS on another instrument, NFA
and ESA are also SNS compatible instruments. Check the device is not
an Keysight power sensor which uses the same cable interface.
The Keysight Smart Noise Source connected to the MXA has failed to be
read by the application. Please disconnect and reconnect the SNS. If
this continues to fail, then the SNS may have had its EEPROM corrupted
or another hardware fault exists. Check SNS on another instrument, NFA
and ESA are also SNS compatible instruments. Check the device is not
an Keysight power sensor which uses the same cable interface.
A parity bit was not correct when the data was received. For example, on
a parallel port.
A stop bit was not detected when data was received. For example, on a
remote bus port.
A software or hardware input buffer on a port overflowed with data
because of improper or nonexistent pacing.
There was a time-out problem in the instrument. The exact problem
cannot be specifically identified.