Instrument Messages
Condition Messages
For example, message 721 indicates that the DC Coupled condition has been
detected; message 1721 indicates that condition has been cleared.
Unless otherwise noted in the Error or Warning column below, these are status
bits only, with no corresponding error message or number
Err#
Bit in
Message
status
register
701
0
Calibrating
703
1
Settling
705
2
unused
707
3
Sweeping
709
4
Measuring
711
5
Waiting for Trigger
713
6
Waiting for Arm
715
7
unused
717
8
Paused
719
9
Source Sweeping
721
10
DC Coupled
723
11
unused
725
12
Source Waiting for
Trigger
727
13
unused
729
14
unused

801 to 899, Temperature

This series of messages corresponds to the bits in the
STATus:QUEStionable:TEMPerature
System" on page
corresponding bit in that register.
Instrument Messages
Error or
Warning
status bit only
status bit only
status bit only
status bit only
status bit only
status bit only
status bit only
status bit only
W
status bit only
register (see
116). The second column in the table shows the
More Information
The "Source Sweeping" bit is used to
indicate various conditions, depending on
the Mode of operation:
In the List Sequencer mode, it is used to
indicate that the sequencer is running
In the EMI Receiver mode, it is used to
indicate that scan results are available
from the Frequency Scan measurement
"X-Series Status Register