Settings Conflict Errors - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-365
Time out error;
Operation on source
is held. If analyzer
is waiting for
trigger, change to
free run to proceed
–221 Settings Conflict Errors
This is one of the messages in the standard SCPI range from
Execution Errors" on page
Err#
Message
-221
De-emphasis only
available in FM
-221
Function not available in
Zero Span
-221
Settings conflict:
Timebase DAC not
available with Pulse
selected
-221
Settings conflict;
-221
Settings conflict; *.CSV
file format is not
available in this
measurement.
-221
Settings conflict; <Q
Param> cannot be changed
when Q same as I
-221
Settings conflict;
<trigger source> trigger
is not available in
current IF Path
Instrument Messages
Verbose/Correction Information
This is a warning message.
This warning can only occur when issuing SCPI to operate the source
while the analyzer is waiting for a trigger. When the analyzer is waiting
for a trigger, source operation is set to pending. (Trigger Source
operation is an exception)
To avoid this, do not operate the source when the analyzer is waiting for
a trigger. If source and analyzer both are utilized, configure the source
parameters ready and run before analyzer waits for trigger.
If both the analyzer and source are waiting for a trigger, the only
operations allowed on the source are trigger source operations, like
external, manual, or bus trigger.
67.
Verbose/Correction Information
The de-emphasis function is only available if FM demod is
selected.
The function you are trying to access is not available in zero span.
The manual adjustment of the Timebase DAC cannot be
performed when the currently selected timebase is Pulse.
A legal command was received but it could not be executed due
to the current device state.
You cannot load or save base instrument traces, as this is not
supported by the Log Plot measurement.
When the "Q Same as I" parameter is set to Yes, the I parameter
value is copied to <Q Param> and the <Q Param> value cannot be
changed. Set Q Same as I to No to enable explicit control of the
<Q Param> value.
The trigger source (Video, RF Burst, I/Q Mag, etc.) is not available
when using the current IF Path, for example FMT is not available
in the 1 GHz IF Path
"–200 to –299,
48