To 499, Power - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Condition Messages
Err#
Status
Message
Register
Bit
315
7
Preamp: Accy unspec'd
<100 kHz
315
7
Preamp: Accy unspec'd
<10 MHz
315
7
Preamp: Accy unspec'd
<30 MHz
317
8
unused
319
9
unused
321
10
unused
323
11
unused
325
12
unused
327
13
unused
329
14
unused

401 to 499, Power

This series of errors corresponds to the bits in the
STATus:QUEStionable:POWer
on page
that register.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These error numbers can be seen in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 409 indicates a 50 MHz Oscillator Unleveled condition has
been detected, error 1409 indicates that failure has been cleared.
Instrument Messages
register (see
116). The second column in the table shows the corresponding bit in
Error or
Verbose/Correction Information
Warning
W
You are trying to use the Internal Preamp at
a frequency below its specified range. The
Internal Preamp will function at lower
frequencies, but the performance is not
specified below 100 kHz.
W
You are trying to use the USB Preamp at a
frequency below its specified range. The
USB Preamp will function at lower
frequencies, but the performance is not
specified below 10 MHz.
W
You are trying to use the LNA at a frequency
below its specified range. The LNA will
function at lower frequencies, but the
performance is not specified below 30 MHz
in PXE.
W
W
W
W
W
W
W
"X-Series Status Register System"
106