Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict;
Electronic attenuator
unavailable with Preamp
on
-221
Settings conflict; EMI
Avg and Voltage Avg cannot
be selected
simultaneously
-221
Settings conflict; EMI
Avg/RMS Avg and Average
detector can't be used
together
-221
Settings conflict; EMI
Detectors are not
available in FFT sweep
-221
Settings conflict;
Enabled modulation source
conflicts with previous
modulation source.
Previous modulation
disabled
-221
Settings conflict;
External Mixer not
available
-221
Settings conflict; FAST
method can only be used
while Radio Std is W-CDMA
-221
Settings conflict;
Feature not available for
Option <abc>
-221
Settings conflict;
Feature not available in
this View
-221
Settings conflict;
Feature not supported for
selected source
-221
Settings conflict;
Feature not supported for
this Input.
Instrument Messages
Verbose/Correction Information
The internal preamp is on. Electronic attenuator cannot be used
while you are using the internal preamp.
You cannot turn on both EMI Avg and Voltage Avg detectors
concurrently. If the EMI Average detector is selected, then
Voltage Average detector cannot be turned on.
User is not allowed to turn on EMI Avg/RMS Avg and Average
detector together. They are always mutually exclusive.
QPD, EMI Average, EMI Peak, and MIL Peak are not allowed when
in the manually selected FFT sweep mode.
Turning one modulation format (like AM, PM, FM or ARB) on when
another modulation format is already on results in the previous
modulation format being turned off and the generation of an
error.
A command has been sent to reference the External Mixer in a
model that does not contain it
This functionality is not part of the instrument configured with the
indicated option
Some functionality is available in one View, but not in another.
(See the Views under the View/Display key.) This error occurs if
you send a SCPI command or push a gray-out key that is not
available in the current selected View.
You have asked for a feature that the selected source is not
capable of.
Some functionality is not available when certain Inputs are
selected. For example, Trigger Holdoff is not available for the
BBIQ input