Hardware Error - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-232
Invalid
format;<filename>
contains invalid
waveform header and
cannot be loaded into
ARB memory
-232
Invalid
format;<filename> has
less than 500 IQ
samples
-232
Invalid format;Syntax
error on source step
<step number>
-232
Invalid format;Unable
to open encrypted
waveform file
<filename>
-233
Invalid version
-240

Hardware error

-240
Hardware error; See
details in Windows
Event Log under SA
-241
Hardware missing
-241
Hardware missing;
Input not available
-241
Hardware missing;
Internal preamp not
available at all
frequency points
-241
Hardware missing; not
available for this
model number
-241
Hardware missing;
Option not installed
-241
Hardware missing;
Output not available
Instrument Messages
Verbose/Correction Information
An attempt was made to load, into ARB memory, a Signal Studio
waveform file that contains an invalid waveform header.
An attempt was made to load, into ARB memory, a waveform file, but
the file does not contain enough IQ samples.
Load an ARB waveform with a sample number equal to or greater than
500.
An attempt was made to load in a list sequencer file that was
incorrectly formatted.
An attempt was made to load, into ARB memory, a Signal Studio
waveform file, but the file cannot be opened due to invalid file format.
A legal data element was found but could not be used because the
version of the data is incorrect. For example, state data changes as
new instrument features are added, so old state files may not work in
an instrument with a newer version of software.
A legal program command or query could not be executed because of
a hardware error. The exact problem cannot be specifically identified.
The internal data acquisition system detected a problem at startup and
logged the details in the Windows Event Log.
The operation could not be performed because of missing hardware;
perhaps the optional hardware is not installed.
The hardware required is not part of this model or the option is not
installed
The Internal Preamp is currently turned on, but the measurement is
being performed completely or partially outside the range of the
preamp. It is recommended that the user turns preamp off to ensure
consistent results across the entire measurement.
The hardware required is not part of this model
The optional hardware is not installed.
The hardware required is not part of this model or the option is not
installed
77