Mass Storage Error; File <Filename> Wrong Type - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-250
Mass storage error;
File <filename> wrong
type
-250
Mass storage error;
File <filename>. Model
and serial number
fields don't match
selected sensor
-250
Mass storage error;
File contains
incorrect data for
this operation
-250
Mass storage error;
File empty
-250
Mass storage error;
Invalid register
number for *SAV or
*RCL Mass Storage
error
-250
Mass storage error;
Lock violation
-250
Mass storage error;
Mkr Table must be on
to save Mkr Table as
Meas Results
-250
Mass storage error; No
file names available
-250
Mass storage error;
Open failed
-250
Mass storage error; Pk
Table must be on to
save Pk Table as Meas
Results
-250
Mass storage error;
Read fault
-250
Mass storage error;
Register <number>
empty
Instrument Messages
Verbose/Correction Information
Attempt to import a data file that is not the proper type for this
operation.
The model and serial number fields don't match the selected sensor.
Check cal factor .xml file content.
There is a mismatch between the file data type of the file specified and
the destination indicated. For example, a correction set cannot be
loaded/imported into a limit line.
Cannot save trace because it contains no data. Check that the trace is
turned on and contains some valid data.
You have used the *SAV command to save a state to a non existent
state register.
Or
You have used the *RCL command to recall a state register that wasn't
previously saved with the *SAV command.
The process cannot access the file because another process has
locked a portion of the file.
You have to have a Marker Table on the screen before you can save it.
Turn on the Marker Table and try again.
Attempt to use the auto file name generation when all 10,000 file
names are taken.
The system cannot open the device or file specified. This could be
because the storage media is full, or possibly due to a filename error. If
using an external storage devise, check that the devise is properly
formatted.
You have to have a Peak Table on the screen before you can save it.
Turn on the Peak Table and try again.
The system cannot read from the specified device.
Attempt to recall a register with nothing in it
79