SA (Spectrum Analyzer) Mode
SA Mode Settings

Video Bandwidth (Video BW)

Video BW is a smoothing operation that is performed after measurement data
is acquired. The trace data is effectively smoothed so that the average power
level of the displayed noise is the same, but the peaks and valleys of adjacent
data points are smoothed together. More smoothing occurs as the Video BW is
set lower. However, as the Video BW is narrowed, the sweep speed becomes
slower.
How to set VBW
— Press BW 2.
— Then
— Then choose from the following:
To change this setting from Man to Auto, press Video BW twice.
The current Video BW setting is shown at the bottom of the screen.
# VBW x.xx XHz (#) means manual setting.
When the Res BW/Video BW ratio exceeds 10,000, a
may appear to indicate that the Video BW filter has reached the maximum
capacity for averaging.

Sweep Type

In SA mode, the FieldFox uses two sweep types to process inputs signals. The
sweep type that is currently being used (FFT or Step) is displayed in the
lower-left corner of the FieldFox screen.
For a more comprehensive tutorial, see Spectrum Analysis Basics (App Note
150) at
How to set Sweep Type
— Press BW 2.
— Then
— Then
— Then choose from the following:
164
Video BW
—Auto
Couples Video BW to the ResBW in a 1-to-1 ratio.
—Man
Enter a Video BW value using the numeric keypad, the
arrows, or the rotary knob. Then press a multiplier if necessary or
press Enter. Setting VBW higher than RBW may smooth the trace.
.
http://literature.cdn.keysight.com/litweb/pdf/5952-0292.pdf