Instrument Messages

Advisory Messages

Advisory Messages
An advisory is simply a message that lets you know something useful—for
example "File saved successfully" or "Measuring fundamental." Operation
completion and running status indications are common types of Advisories.
Advisories have no number and are not logged in the error queue. They are not
defined in the
Advisories include gray-out "settings conflict" errors. These gray-outs are
benign (that is, changing them has no impact on the current measurement).
Advisories are Event-type errors only. They are never Conditions.
Message
All Auto/Man functions have
been set to Auto
All Auto/Man functions have
been set to Auto.
Allowable Center Frequency
exceeded for the current
span
Allowable Span exceeded for
the current center
frequency
Allowable Start Frequency
exceeded for the current
span
Allowable Stop frequency
exceeded for the current
span
Allowable X Ref Level
exceeded for the current
Capture Length
Allowable X Scale/Div
exceeded for the current
Capture Length
Already in Single, press
Restart to initiate a new
sweep or sequence
Auto sweep time rules do
not apply in FFT sweeps
Instrument Messages
1999 SCPI Syntax & Style
Description/Correction Information
Message generated by pressing the Auto Couple front-panel key.
When rotating the knob or step up/down keys to change the Center frequency, the
value of the Span is kept constant. Therefore, the center frequency is limited by the
frequency range of the instrument.
When rotating the knob or step up/down keys to change the Span, the value of the
Center frequency is kept constant. Therefore, the span is limited by the frequency
range of the instrument.
When rotating the knob or step up/down keys to change the Start frequency, the
value of the Span is kept constant. Therefore, the start frequency is limited by the
frequency range of the instrument.
When rotating the knob or step up/down keys to change the Stop frequency, the
value of the Span is kept constant. Therefore, the stop frequency is limited by the
frequency range of the instrument.
The X Ref Level and X Scale/Div is limited by the Capture Length of the
measurement.
The X Ref Level and X Scale/Div is limited by the Capture Length of the
measurement.
The instrument is already in the single state. If you want to start a new sweep or
sequence, press the Restart key instead.
FFT sweeps do not use the auto sweep time rules, so the rules setting cannot be
changed from the front panel. The setting can be changed remotely and it will have
no effect on the current operation unless the analyzer is switched out of FFT
sweeps.
Standard.
36