Antenna Unit Already In Used In Another Range - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-241
Hardware missing;
Power sensor hardware
is missing
-250
Mass storage error;
-250
Mass storage error;
Access denied
-250
Mass storage error;
Bad path name
-250
Mass storage error;
Can only import single
trace .csv files
-250
Mass storage error;
Can only load an
Antenna Unit into
Correction 1
-250
Mass storage error;
Can only load one file
with Antenna Unit per
range
-250
Mass storage error;
Cannot make
-250
Mass storage error;
Different
Antenna Unit already
in used in another
range
-250
Mass storage error;
Different Antenna Unit
already in use
-250
Mass storage error;
Directory not found
-250
Mass storage error;
Failed to Load trace.
Bad file format.
-250
Mass storage error;
File <filename> and
instrument version
mismatch
Instrument Messages
Verbose/Correction Information
There is no power sensor detected at the selected power meter. You
need to setup the power sensor before proceed.
A problem was found with the mass storage device (memory, disk
drive, etc.). The exact problem cannot be specifically identified.
Access is denied.
The specified path is invalid.
Trace files containing multiple traces can not be imported. However, if
you need to recall multiple traces you can use the Save and Recall
functions rather than the Import and Export functions.
The only Correction register that supports Antenna Units is number 1.
You have attempted to load an Ampcor file which contains antenna
units into another register
You already have a Correction file with Antenna Unit in this range.
Attempt to import more than one Corrections file with Antenna Unit is
not allowed.
The directory or file cannot be created.
You have attempted to import a Corrections file with Antenna Unit
different from another range.
Attempt to import Corrections file with Antenna Unit that differs from
an in-use correction.
The system cannot find the path specified.
The load trace operation could not be completed, as the input file was
not in the expected format. You can only load traces that were
previously saved using the 'Save Trace' feature.
While opening a file, there was a mismatch between file version or
model number with instrument version or model number. The import
still tried to load as much as possible, but you should check it closely.
78

Hide quick links:

Advertisement

Table of Contents
loading