Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict;Range
<{0}> is turned off as
total range points >
4,000,001.
-221
Settings conflict;Scan
Time & Points do not apply
in current Step/Time
Control.
-221
Settings conflict;Screen
limit reached
-221
Settings conflict;Screen
SCPI cannot be used when
Display is disabled
-221
Settings conflict;Step
Size & Meas Time do not
apply in current
Step/Time Control.
-221
Settings conflict;The
number of input
parameters is too large
and is truncated to
current list step number
-221
Settings
conflict;Transition time
may be short on step <step
number>.
–200 to –299, Execution Errors
This section lists all messages in the range except –221 messages. For details
of –221 messages, see
Note that Execution Errors are divided into subclasses, as follows:
–21x
–22x
–23x
–24x
–25x
Instrument Messages
Verbose/Correction Information
Max of Total range points is 4,000,001. Reduce Scan Points or
increase Step Size in order to turn on that range.
The Scan Time & Points are not available when Step/Time Control
is set to Step & Dwell.
You requested to add another screen with the number of screens
at the maximum.
The instrument is operating with display disabled
(DISPlay:ENABle OFF), you cannot create new Screens with the
display disabled.
The Step Size & Meas Time are not available when Step/Time
Control is set to Scan Time&Pts.
An attempt to set list sequence parameter whose index exceeds
the "Number of Steps". This way only those parameters whose
indexes fall in number of steps will be accepted.
Reduce the number of parameters to be no large than "Number of
Steps".
An attempt was made to set the source step transition time to a
value shorter than the hardware settling time
This is provide as a warning. Instrument operation may not be
impacted because the hardware could settle more quickly than is
normal. You can eliminate this warning by setting a longer
transition time.
"–221 Settings Conflict Errors" on page
Trigger errors
Parameter error
Data corrupt or stale (invalid data)
Hardware error
Mass storage error
48.