Settings Conflict; Averaged Acquisition Duration Time Exceeds Maximum Value, Acq - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict;
Averaged acquisition
duration time exceeds
maximum value, Acq=#
-221
Settings conflict; Basic
Discrete PAvT measurement
cannot be averaged, Acq=#
Step=#
-221
Settings conflict; Basic
IQ data measurement
cannot be averaged, Acq=#
Step=#
-221
Settings conflict; BTS
gain is not available in
this Mode
-221
Settings conflict; Cal
only available when
Source Mode is Tracking
-221
Settings conflict;
Calibrate TRFL is only
available in Manual Range
Switching when Recal is
needed
-221
Settings conflict;
Calibration cannot be
performed without valid
ENR data
-221
Settings conflict;
Cancellation is not
available while measuring
DANL floor.
-221
Settings conflict;
Cancellation Ref trace
has no data.
-221
Settings conflict;
Cancellation trace has
different X-Scale
Instrument Messages
Verbose/Correction Information
Total acquisition duration time after averaging (average count *
acquisition duration) exceeds maximum value.
Reduce the average count or decrease the acquisition duration
time.
The average count of acquisition which contains basic Discrete
PAvT measurement cannot be greater than 1.
Set the average count of current acquisition to 1. Or remove the
Basic Discrete PAvT measurement from this analysis step.
The average count of acquisition which contains basic IQ data
measurement cannot be greater than 1.
Set the average count of current acquisition to 1. Or remove the
Basic IQ data measurement from this analysis step.
Base Transceiver Station gain correction is not available in some
Modes, or in some measurements (for example, the SA
measurement).
You must be in Tracking Source mode to use the Cal functions
under Normalize. Press Source, Source Mode and set it to
Tracking.
You can only perform this operation when in Manual Range
Switching mode and input power fall in the recal range.
The cal ENR table has no values in it, and hence the cal cannot be
performed. Correct by either populating the cal ENR table, set
ENR mode to Spot, or set the 'Use Meas Table Data for Cal' to
'On' .
Phase Noise cancellation does not make sense when measuring
the DANL Floor, so for this reason it has been disabled
When performing phase noise cancellation, you need to supply a
reference trace that will be used to cancel out the background
noise of the analyzer. The reference trace must be in Reference
(View) mode, and selected by the Ref Trace parameter under the
Cancellation menu
Reference trace for the cancellation has a different range of
X-axis against the target trace
50