To 95, Calibration Failure (Extended) - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Condition Messages
Msg#
Bit in
Message
status
register
72
4
Input Attenuation not
calibrated
74
5
unused
76
6
unused
78
7
unused
80
8
MPA Align required
82
9
unused
84
10
unused
86
11
Characterize Preselector
required
88
12
Characterize Noise Floor
required
90
13
unused
92
14
unused

67 to 95, Calibration Failure (Extended)

This series of errors corresponds to the bits in the
STATus:QUEStionable:CALibration:EXTended:FAILure
"X-Series Status Register System" on page
table shows the corresponding bit in that register.
An event with the error number shown in the table means the condition has
been detected. When the condition is cleared, an event with the error number
plus 1000 is generated. These numbers can be viewed in the Show Errors
screen, along with the DETECTED and CLEARED indicators.
For example, error 71 indicates that the Characterize Preselector Failure has
been detected, error 1071 indicates that failure has been cleared.
This register is summarized as bit 9 of the
register, as described in the section
page
Err#
Bit in
Message
status
register
67
0
Align 9kHz to 30MHz
failed
Instrument Messages
88.
Error or
More Information
Warning
E
Corrected measurements have been
requested and the required RF front-end
setting of x dB has not been calibrated.
W
EXT only
W
Align
W
116). The second column in the
STATus:QUEStionable:CALibration
"36 to 64, Calibration Needed or Failed" on
Error or
More Information
Warning
W
On PSA, this was error 13749
sub-register (see
91