Instrument Messages
Condition Messages
An event with the number shown in the table means the condition has been
detected. When the condition is cleared, an event with the number plus 1000
is generated. These error numbers can be viewed in the Show Errors screen,
along with the DETECTED and CLEARED indicators.
For example, message 801 indicates that the Ref Osc Oven Cold condition has
been detected; message 1801 indicates that condition has been cleared.
This register is summarized as bit 4 of the
described in the section
Err#
Bit in
Message
status
register
801
0
Reference Oscillator Oven
Cold
803
1
Over temperature
803
1
Over temperature: let cool
and reboot
803
1
Over temperature; T/R port
805
2

Over current: reboot

807
3
unused
809
4
unused
811
5
unused
813
6
unused
815
7
unused
817
8
unused
819
9
unused
821
10
unused
823
11
unused
825
12
unused
Instrument Messages
STATus:QUEStionable
"601 to 699, Error Summaries" on page
Error or
More Information
Warning
W
Reference Oscillator oven has not yet
reached the correct temperature for the
instrument to meet specifications.
E
E
The internal temperature of the
instrument exceeds operating levels.
The LO or the IF has signaled an
overtemperature condition. Ensure
proper airflow and reboot the
instrument. See the Windows Event Log
under SA for details.
E
High temperature detected, T/R input
switched to open state, PXIe Chassis
may power down.
E
The IF has signaled an over current
condition. Reboot the instrument. See
the Windows Event Log under SA for
details.
register, as
109.
112