Instrument Messages
Condition Messages
This register is summarized as bit 3 of the
described in the section
Err#
Bit in
Message
status
register
401
0
RPP Tripped
403
1
Source Unleveled
405
2
Source LO Unleveled
407
3
LO Unleveled
409
4
411
5
413
6
Input Overload
413
6
Input Overload;T/R port
413
6
Input Overload;Gen reverse
power
415
7
417
8
419
9
Full Bypass Enabled, maximum
safe input power reduced
421
10
Input caution
421
10
Input caution;T/R unprotected
423
11
unused
425
12
unused
427
13
unused
Instrument Messages
STATus:QUEStionable
"601 to 699, Error Summaries" on page
Error or
Warning
W
W
E
E
E
E
W
W
W
register, as
109.
More Information
(not currently in use)
(not currently in use)
(not currently in use)
The maximum input power at the T/R
port has been exceeded. Accurate
measurements cannot be made, please
reduce the input level
Reverse power detected at the Gen
output port. Please remove the reverse
power.
Full Bypass Enable is selected, which
means that when the Low Noise Path
switches in (when Start Freq >3.6 GHz
and the Preamp is either not licensed,
set to Low Band or Off) and Attenuation
is set to 0 dB, there will be a direct AC
connection between the input and the
first converter, which puts the first
converter at high risk to be damaged
by high AC power. Keep the power at
the input below +20 dBm.
User is cautioned that in this mode,
high power at the T/R input can cause
damage to the input circuitry