Settings Conflict; <Trigger Source> Trigger Is Not Available While Input Is <Input Port - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict;
<trigger source> trigger
is not available while
input is <input port>
-221
Settings conflict; A
Valid User Cal is
required. Optimize
aborted
-221
Settings conflict;
Acquisition # Analysis
Step # not exist
-221
Settings conflict;
Acquisition # not exist.
-221
Settings conflict;
Acquisition duration time
too short, Acq=#
-221
Settings conflict;
Analysis Interval should
be less than ** ms, Acq=#
Step=#
-221
Settings conflict;
Analysis Offset should be
greater than # us, Acq=#
Step=#
-221
Settings conflict; At
least one Phase must be
turned on
-221
Settings conflict; At
least one Protective
Earth must be turned on
-221
Settings conflict; Auto
Carrier Frequency is only
available when Modulation
Mode is FM or PM
-221
Settings conflict; Auto
Set RF Level failed, Too
much dynamic range is
requested
-221
Settings conflict; Auto
Tune not available in
Tracking Source mode
Instrument Messages
Verbose/Correction Information
The trigger source (Video, RF Burst, I/Q Mag, etc.) is not available
with the current input port (RF, IQ, etc.)
Optimize Preselector can only be performed if a valid user cal
exists and is applied to current results. Perform a user cal first or
apply existing cal.
The configured Analysis Step does not exist.
Add the specified analysis step into the sequence.
The configured acquisition does not exist.
Add the specified acquisition into this sequence.
Acquisition duration cannot be less than < (analysis offset +
analysis interval).
Increase the acquisition duration time or decrease the analysis
offset and analysis interval.
The analysis interval is too large for basic transmit power
measurement.
Decrease the analysis interval.
The Analysis Offset should be greater than # µs due to physics
requirement.
Increase the time of Analysis Offset.
At least one Protective Earth must be turned on
You are not allowed to change the Auto Carrier Frequency
settings in AM Modulation Mode.
The sequence requests too much dynamic range, auto range
cannot be preformed correctly.
Change the Peak Power or DUT Expected Power.
The Auto Tune feature cannot be used when you are using a
Tracking Source.
49