Instrument Messages
Condition Messages
Err#
Status
Message
Register
Bit
309
4
User Cal; Freq outside cal
range
309
4
User Cal; Cal will be
interpolated
309
4
User Cal; Adjusted for new
RBW
311
5

Calibration

311
5
Calibration; ENR table
extrapolated
311
5
Calibration; No ENR data
present
313
6
Source Uncal
313
6
Source Uncal;adj Start Freq
or RBW
313
6
Source Uncal;adjust Source
Amplitude
315
7
Preamp: Accy unspec'd
<XX kHz
Instrument Messages
Error or
Verbose/Correction Information
Warning
E
The existing user cal has been invalidated
because the current measurement
frequencies lie partially or wholly outside
the range of frequencies used for user-cal.
(UNCAL)
W
The measurement frequency range has
been changed such that it is a subset of the
calibrated range. (~CAL)
W
The measurement RBW has been changed
since the last calibration (~CAL)
W
W
One or more calibration or measurement
frequency points exceed the currently
loaded Cal or Meas ENR Table frequency
ranges. The corresponding ENR table's
lowest frequency ENR value will be re-used
for frequencies less than the table range,
and the highest frequency ENR value will be
re-used for frequencies greater than the
table range.
(~ENR)
W
No ENR Data (ENR)
W
W
While using a Tracking Source, you must
make sure the Start Frequency is high
enough to avoid capturing LO feedthrough
in the trace. This depends on both Start
Freq and RBW. If you get this message,
increase your Start Freq or narrow your
RBW.
W
While using a Tracking Source, source
amplitude can be set at specific
un-calibrated range, but the performance is
not specified. If you get this message, adjust
source amplitude.
W
Preamp will function at lower frequencies,
but the performance is not specified below
XX kHz (XX is model number specific).