Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; Step
keys are not available to
modify this function
-221
Settings conflict; Sweep
Setup only available in
swept measurements
-221
Settings conflict; Sweep
Time cannot be
auto-coupled in FFT
sweeps
-221
Settings conflict; Sweep
Time cannot be
auto-coupled while in
Zero Span
-221
Settings conflict; Sweep
Time cannot be set while
in FFT sweeps
-221
Settings conflict; Swept
IF Gain High not available
when FFT IF Gain = Manual
Low
-221
Settings conflict; Swept
LO not available when freq
mode is Fixed
-221
Settings conflict; Swept
Type=Swept is not
available while in Gated
FFT
-221
Settings conflict; System
Display Settings,
Annotation is Off
-221
Settings conflict; T hot
must be greater than T
cold
-221
Settings conflict; The
parameter cannot be
changed in FAST mode
-221
Settings conflict; Time
Domain Scan is not
available with EMC "None"
Standard
Instrument Messages
Verbose/Correction Information
You should select a specific value for this function. So using the
Up/Down step keys to scroll through values is not allowed.
The current measurement uses FFT mode and so does not use the
Sweep Setup menu
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
You cannot send the remote command to set the sweep time to
auto while you are in zero span.
The sweep time for FFT sweeps is set by the calculations. It
cannot be manually controlled.
When FFT IF Gain is manually set to Low, you cannot set the
Swept IF Gain to High because that would make the Reference
Level couplings wrong in swept mode.
SCPI only message. The LO Mode cannot be set to Swept when
the freq mode is set to fixed. Change the freq mode away from
fixed, or perform the measurement at several fixed frequencies.
If you have selected gated FFT then you are using the FFT sweep
type and you cannot select the swept type of sweeping.
This is an override that turns off many of the annotations. This is
available as a security feature.
The Tcold value set under Meas Setup/ENR/Tcold, needs to be
lower than the Thot value currently being set. Tcold is often taken
as the ambient temperature of the noise source. If using an SNS
the Tcold value may be read automatically before every sweep.
TDS only supports CISPR and MIL standards