Instrument Messages
Event Messages
Err#
Message
-200
No Peak Added to
Signal List
-200
No unmarked signals
-200
Signal List is Empty
-200
Signal List is Full
-200
Signal Selected is not
in the Signal List
-200
Signal selected is out
of display range
-201
Invalid while in local
-202
Settings lost due to
rtl
-203
Command protected
-203
Command protected;
feature not licensed
-203
Command
protected;Feature Not
Supported. No
appropriate waveform
license installed for
<filename>
-210
Trigger error
-211
Trigger ignored
-212

Arm ignored

-213
Init ignored
-214
Trigger deadlock
Instrument Messages
Verbose/Correction Information
No signal peak was added from the Search to Signal List as there is no
signal peaks found within the defined parameters of the search
criteria.
Cannot perform the selected function because no signal was
unmarked. You must unmark the peak of interest before selecting the
function.
Cannot perform the selected function because the signal list contains
no data.
Cannot perform the selected function because the signal list is full.
Please clear the list.
The signal selected is not the list. You only can perform the operation
on signal that is already available in the list.
Signal selected is out of display range. You must extend the display
range if you wish to view the selected signal on the graph display.
The command cannot be executed while the instrument in Local
control.
A "return to local" control was forced and some settings were lost as a
result of this.
The command could not be executed because it is disabled. It was
disabled by licensing or password protection
The specified feature, for example "N9073A-TR2" is not licensed. The
license may have expired. You cannot use it until you get a license.
An attempt was made to select a waveform while the required license
is not installed
Install necessary Signal Studio waveform licenses or multi-pack
licenses.
A trigger error has occurred, but the exact problem cannot be
specifically identified.
A GET, *TRG or other triggering signal was received, but it was ignored
because of timing considerations. For example, maybe the instrument
was not ready to respond when the command was received.
An arming signal was received, but it was ignored.
An initiate trigger/sweep request was received and ignored, because
another measurement was already in progress.
The trigger source for the initiation of a measurement is set to GET,
and the following measurement query was received. The measurement
cannot be started until a GET is received, but the GET would cause an
INTERRUPTED error.