Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; Input
Port is not available
while occupied by Output
of incompatible frequency
-221
Settings conflict; Input
Port is not available
while port bank is
occupied by Output
-221
Settings conflict; Input
Z unavailable when probe
sensed
-221
Settings conflict;
Instrument Gain LOW is not
supported when frequency
> 3.6 GHz. Auto change it
to ZERO, Acq=#
-221
Settings conflict;
Internal Preamp is not
available for this
measurement
-221
Settings conflict;
Invalid trace number
-221
Settings conflict; Knob
is not available to modify
this function
-221
Settings conflict; Limit
cannot be auto-coupled
while freq is out of
CISPR14 range
-221
Settings conflict; LO
Phase Noise Adj not
available
-221
Settings conflict; LPF
300 kHz setting is not
available in Audio
Measurement
-221
Settings conflict; Manual
RF Input Ranging is not
available
-221
Settings conflict; Marker
1 Trace Update=off turns
off Signal Track
Instrument Messages
Verbose/Correction Information
You attempted to select a port as an Input for a mmWave
Transceiver of a given frequency range, but the Output is already
configured for a mmWave Transceiver of an incompatible
frequency range. First set the Output Port to None and then
select an Input port.
The desired destination port bank for the Input Port is already
occupied by the Output Port. First set the Output to a different
port bank or None. Note the Output must be OFF before it can be
changed to a different port.
A probe is connected and the Input Z is set based on the probe
type. It cannot be changed manually.
When the acquisition frequency is change to be greater than
3.6GHz, the Instrument Gain will be set to ZERO automatically if
the current value is LOW.
This is a warning message, no action is required.
You are not allowed to change internal preamp in the current
measurement.
The subopcode used to specify the trace number is invalid for this
measurement or query
You should select a specific value for this function. So scrolling
through values with the knob is not allowed.
Settings conflict; Limit cannot be auto-coupled while freq is out of
CISPR14 range
For instruments without the Dual-Loop LO, this feature is not
available
You are not allowed to change LPF setting to 300 kHz in Audio
Measurement because the input only can read until 200 kHz.
You are not allowed to change the Input Attenuation in the current
range.
Signal Track not available unless the trace containing Marker 1 is
updating
57