To 599, Frequency - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Condition Messages
Err#
Bit in
Message
status
register
429
14
unused

501 to 599, Frequency

This series of errors corresponds to the bits in the
STATus:QUEStionable:FREQuency
System" on page
corresponding bit in that register.
An event with the number shown in the table means the condition has been
detected. When the condition is cleared, an event with the number plus 1000
is generated. These error numbers can be viewed in the Show Errors screen,
along with the DETECTED and CLEARED indicators.
For example, error 503 indicates a Frequency Reference Unlocked condition
has been detected, error 1503 indicates that failure has been cleared.
This register is summarized as bit 5 of the
described in the section
Err#
Bit in
Message
status
register
501
0
Source Synth Unlocked
503
1
Frequency Reference Unlocked
505
2
nd
2
507
3
unused
509
4
LO Unlocked
511
5
unused
513
6
IF Synthesizer Unlocked
515
7
Calibration Oscillator
Unlocked
517
8
unused
519
9
Demodulation
521
10
Ref missing or out of range
Instrument Messages
116). The second column in the table shows the
"601 to 699, Error Summaries" on page
LO Unlocked
Error or
More Information
Warning
register (see
"X-Series Status Register
STATus:QUEStionable
Error or
More Information
Warning
E
E
E
E
E
E
E
E
register, as
109.