Input Overload;Adc Over Range - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Condition Messages
Err#
Bit in
Message
status
register
141
4

Input Overload;ADC over range

141
4
Input Overload;I/Q ADC over range
141
4
Input Overload;I/Q Voltage over
range
141
4
Input Overload;RF Preselector
Overload
143
5
Over Range
145
6
Under Range
147
7
Insufficient Data
147
7
Insufficient Data; Incr. Demod
Time
147
7
Insufficient Data; frequency list
empty
147
7
Insufficient Data; ENR table
empty
Instrument Messages
Error or
More Information
Warning
E*
The signal at the input to the IF
section is too high. You should
increase the attenuation or lower the
signal level.
E*
The I or Q input exceeds the ADC
upper limit.
E*
The input voltage on the I or Q
channel exceeds the channel limit. In
differential mode the over voltage may
occur without causing an ADC
overload, for example, if I is at +5.01 V
and I-bar is at +5.0 the ADC will be in
range but both I and I-bar will exceed
the voltage limit.
E*
The level at the input of the MXE RF
Preselector has exceeded tolerances,
reduce the input level
Status Bit
The signal at the input for this
Only
measurement  s too high. You should
increase the attenuation or decrease
the signal level. 
Status Bit
The signal at the input for this
Only
measurement is too low. You should
decrease the attenuation or increase
the signal level. 
E
E
There is insufficient acquisition data to
provide accurate metrics. You should
increase the Demod Time to acquire
enough data.
E
A measurement was attempted with
List frequency mode or a SCPI query
of the frequency list table was made
and the frequency list table is empty.
E
A measurement was attempted or a
SCPI query of an ENR table was made
and there were no entries in the
relevant ENR table (Common, Meas or
Cal).
94