To 92, Calibration Needed (Extended) - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Condition Messages
Err#
Bit in status
Message
register
54
9
Extended Align Failure
Summary
56
10
unused
58
11
Align Skipped Sum Summary
60
12
Align Now, RF required
62
13
unused
64
14
Align Now, All required

65 to 92, Calibration Needed (Extended)

This series of messages corresponds to the bits in the
STATus:QUEStionable:CALibration:EXTended:NEEDed
"X-Series Status Register System" on page
table below shows the corresponding bit in that register.
An event with the number shown in the table means the condition has been
detected. When the condition is cleared, an event with the number plus 1000
is generated. These error numbers can be viewed in the Show Errors screen,
along with the DETECTED and CLEARED indicators.
For example, message 72 indicates that the "Input Attenuation not calibrated"
condition has been detected, and message 1072 indicates that failure has
been cleared.
This register is summarized as bit 8 of the
register, as described in the section
page
Msg#
Bit in
Message
status
register
65
0
unused
66
1
Align 9kHz–30MHz required
68
2
Align 30MHz–1GHz required
3
Instrument Messages
88.
Error or
More Information
Warning
status bit
This bit is the summary bit for the
only
STATus:QUEStionable:CALibration:EXT
ended:FAILure sub-register.
status bit
This bit is the summary bit for the
only
STATus:QUEStionable:CALibration:SKIP
ped sub-register.
E
E
On PSA, this was error 64
116). The second column in the
STATus:QUEStionable:CALibration
"36 to 64, Calibration Needed or Failed" on
Error or
More Information
Warning
E
An EMI conducted frequency range
alignment is needed.
E
An EMI radiated frequency range
alignment is needed.
sub-register (see
90