Horiba Scientific FluoroMax-4 Operation Manual page 98

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
c
2
Choose correction when setting up the experi-
ment.
a
b
Choose Preferences, then the Instrument Correction Files icon.
The Instrument Correction Files area should display a correction file
for the Detector (S or R). If not, click the Insert button, and browse for
the desired correction file.
Detector Correction-factor file name
mcorrect
S
xcorrect
R
In the Experiment Setup window, choose the Detectors icon
to display the detectors' parameters.
Click the Correction checkbox for the detector you want corrected.
In the Signal Algebra area, a signal with appended "c" appears, denot-
ing a corrected signal:
5-16
Optimizing Data