FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jun 2012)
Analyze the data.
1
Using the
HJY_simple_math
window under the
Analysis/HJY
Tools menu, di-
vide the irrad
file by the
stdlamp file:
The irrad file is the irradiance file associated with the detector.
The new graph now displays the correction factor.
2
Normalize the new mcorrect file.
a
Note:
Choose the correct irradiance file associated with the de-
sired detector.
Using the HJY_normalize window under the Analysis/HJY Tools
menu, divide the irrad file by this minimum signal intensity, a
constant.
8-7
Producing Correction Factors