FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Replace the lamp-housing cover.
Replace the FluoroMax
The FluoroMax
This should confirm that the instrument is roughly calibrated.
4
Open Real Time Control.
a
b
c
d
5
Adjust the xenon lamp's position.
There are three adjustments to opti-
mize the lamp position:
Adjustment
Focus
Tilt and Center
Height
a
b
Yes
No further adjust-
ment is necessary.
®
-4 cover.
STOP.
®
-4 is ready for operation.
Move the excitation monochromator to 350 nm, and the emission
monochromator to 397 nm.
Set the slits to 5-nm bandpass.
Open the excitation shutter and emission shutter.
Set the integration time to 0.5 s.
Allen key re-
quired
3/32"
9/64"
1/8"
Slowly work each adjustment
to optimize the signal.
At maximum signal (>
400 000 cps, depending upon
the spectrometer), the lamp is
optimized.
Remove the Allen keys.
Is the peak
intensity ≥
400 000 cps?
Adjust the lamp:
Continue with the
procedure below.
Focus
Height
6-9
Maintenance
No
Tilt
and
Center