Automated Polarizers
Make sure to use the appropriate Signal in the Signal Algebra area. Add >> it to the
Formulas table. You can add the detector signals at any of five polarization-angle
combinations (VV, VH, HV, HH, VM) to the Formulas list. The polarization and ani-
sotropy ratios are calculated automatically in the Anisotropy-type experiments, and
added by default to the Formulas list.
Click the G factor checkbox if you know the G factor, then enter the G factor in the
box. If you do not know the G factor, leave the G factor checkbox unchecked, and en-
ter 1. The G factor will be measured during the experiment, and automatically used in
the anisotropy or polarization ratios.
Click the Run button when ready.