During acquisition
Data can be acquired either as raw data or as corrected data. A spectrum composed of
raw data exhibits the effects of system parameters, while a corrected spectrum displays
only the properties related to the sample.
Note:
mends subtracting the dark counts, and the spectrum of the blank,
from the data. See the on-line FluorEssence™ help files for specific
instructions.
1
Be sure the instrument configuration has a lay-
out that includes a correction file associated with
the appropriate detector.
a
b
Before applying correction factors, HORIBA Scientific recom-
Caution:
The mcorrect and xcorrect files are custom-
generated at the factory for each particular instrument, and
cannot be swapped.
Choose
Collect in the
toolbar.
A drop-down
menu appears.
Choose
Advanced
Setup, then
System
Configuration.
The System
Configuration window appears:
5-15
Optimizing Data