The following lamp-scan spectrum shows poor resolution in the area around the peak.
0.4473
0.3355
0.2236
0.1118
0
300
Poor lamp scan of 150-W Xe lamp. Note low resolution in the area near the 467-nm
This lack of spectral resolution appears because the slit widths are set too wide. To re-
solve this problem, narrow the slit-widths.
400
500
Wavelength (nm)
peak.
7-5
Xenon-lamp peaks
are unresolved
600
700