Fl4-1011 Four-Position Thermostatted Cell Holder - Horiba Scientific FluoroMax-4 Operation Manual

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
FL4-1011 Four-Position Thermostatted
Cell Holder
The FL4-1011 Four-Position Thermostatted Cell Holder keeps a sample at a constant
temperature from –20°C to +80°C. The temperature is maintained by an ethylene-
glycol–water mixture pumped through from an external circulating temperature bath
(not included). The holder also includes a magnetic stirrer, for mixing turbid or viscous
samples. Also required is the FM-2003 Sample Compartment Accessory.
Caution:
Safety Data Sheet (MSDS) for
information on the hazards of
an ethylene-glycol–water mix-
ture.
Installation
1
Remove the sample-compartment gap-bed.
2
Position the FL4-1011 gap-bed drawer.
3
Tighten with four screws.
4
Attach the ¼" tubing to the brass inlets on the
bottom of the holder.
FL4-1011 Four-Position
Thermostatted Cell Holder.
Refer to your Material
Caution:
may result in flooding and damage to the optics
and electronics of the instrument.
Failure to clamp these hoses securely