Horiba Scientific FluoroMax-4 Operation Manual page 135

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jun 2012)
Calculating emission correction factors
Set up the F-3026 accessory.
1
Attach the correction-
factor light-source to its
gap-bed with the brass
thumbscrews.
Use this orientation. Be sure the power
cable doesn't block optical apertures.
2
Plug the power cable
into the front of the gap-
bed.
3
Remove the standard
sample-compartment
gap-bed, and insert the
correction-factor kit
gap-bed into the sam-
ple compartment.
Fix the gap-bed firmly to the sample compartment with screws through the front
panel.
Run the scan.
1
Start FluorEssence™
and initialize the instru-
ment.
Turn the xenon lamp OFF, and the lamp
for the correction-factor kit ON.
Note:
The source/integrating-sphere module may arrive
premounted atop the special gap-bed.
8-5
Producing Correction Factors