Horiba Scientific FluoroMax-4 Operation Manual page 110

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Adjusting the new xenon lamp
Choices
There are two choices after lamp installation:
Let the lamp "burn in", i.e., run, for 6 h before adjustment of its position.
Set the coarse lamp adjustments immediately. After a 6-h burn-in, set the fine ad-
justments.
For lamp replacement, the major adjustment is to optimize the height screw that was
loosened to remove the old xenon lamp.
Procedure
1
With the lamp
cover still re-
moved, turn on
the FluoroMax
4.
2
Insert a clean cuvette filled with de-ionized wa-
ter in the sample compartment.
3
With the room lights off and ambient stray light
minimized, acquire a xenon-lamp scan and wa-
ter-Raman scan, as explained in Chapter 3.
®
-
Caution:
Intense ultraviolet, visible, or infrared light
may be present when the lamp cover is open, so
wear eye- and skin-protection, such as light-
protective goggles and light-blocking clothing.
6-8
Maintenance