Problem
Poor polarization data
Low polarization ratio
Autopolarizers do not
initialize (they do not
move during initializa-
tion).
Software failure initial-
izing autopolarizers
Cause
Possible Remedy
Improper sample
Adjust sample concentration.
concentration
Photomultiplier
Check that sample signals are in
saturated; slits im-
linear region (< 2 × 10
10 μA on R). Reset slits.
properly set
Dirty cuvette
Clean the cuvette.
Polarizer misa-
Check polarizer alignment.
ligned
Using plastic or cy-
Use rectangular, glass or quartz cu-
lindrical cuvette
vette.
System misaligned
Check system alignment in a gener-
ic layout. Run lamp scan and wa-
ter-Raman scan to check calibra-
tion.
Highly concentrat-
Check Ludox
ed standard
tration: higher concentrations can
cause inner-filter effect, lowering
ratio.
Improperly set slits
Set slits for ~ 1 × 10
Signals much less than this give
excessive contribution from dark
noise, while signals > 2 × 10
are in non-linear region.
System misaligned
Check system alignment in generic
layout. Run lamp scan and water-
Raman scan to check calibration.
Wrong instrument
Check that a configuration with
configuration is
autopolarizers is loaded.
loaded
Wrong instrument
Check that a configuration with
configuration is
autopolarizers is loaded.
loaded
Bad cable connec-
With the system power off, recheck
tions
cable connections.
Computer hang-up
Exit the software, and reboot the
system and host computer.
10-20
Automated Polarizers
6
cps on S, <
®
or glycogen concen-
6
cps in VV.
6
cps