Horiba Scientific FluoroMax-4 Operation Manual page 170

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Re-alignment of polarizers
®
FluoroMax
-4 autopolarizers may be
aligned using a software routine called Po-
larizer Alignment in Experiment Setup.
Using Polarizer Alignment
This routine automatically calibrates autopolarizers. Use a sample of Ludox
gen to run the alignment routine. The software rotates the polarizers in 1° increments
and locates the optimal positions for each autopolarizer. After completion, the anisotro-
py for the scattering solution is measured and displayed for user approval of the align-
ment. If approved, the new calibration positions are saved in the sample-compartment
initialization file, and a log file, POLAR.LOG, is saved with the results of the calibra-
tion procedure. Otherwise, the previous calibration positions are still used.
1
Start FluorEssence™.
2
Open the Experiment Setup window.
3
Click the Accessories icon.
4
Click the Advanced... button.
This opens the Polarizer Alignment window:
Note:
Polarizers are aligned
at the factory, and should
not need realignment.
10-10
Automated Polarizers
®