FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
4
Click the Detectors icon.
This shows the parameters related to detectors, including the G factor, in the
Polarization area.
5
Click the G Factor checkbox to include a G fac-
tor in your measurements.
Activate the checkbox ONLY if you know the G factor. De-activate the check-
box if you want to measure the G factor.
6
Enter a value for the G factor in the field if you
know and want to use a pre-determined value.
Note:
rather than measure it automatically. This
may improve the S/N.
Note:
For detailed information on the G factor, see the Polariz-
ers Operation Manual.
For weak signals, enter the G factor,
5-6
Optimizing Data