Horiba Scientific FluoroMax-4 Operation Manual page 196

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Use
1
Place the sample in a 10 mm × 10 mm cuvette
and insert a magnetic stirring bar.
The stirring bar is available from Bel-Art Products, Pequannock, NJ.
2
Place a cuvette in each holder.
3
Allow the samples to reach the desired tempera-
ture.
4
Turn on the magnetic stirrer.
5
Select the appropriate mixing speed.
The speed at which the sample should be mixed depends on the viscosity of the
sample.
6
Run your experiment as usual.
7
Place the next cuvette in the sample position by
lifting up the knob and rotating the holder.
Be sure to press down, to lock the cuvette into the proper position.
Note:
While the four-position model maintains the temperature
of all four samples, only one sample is mixed at a time.
Note:
Selecting too high a speed may create a vortex, which
could affect the reproducibility of the measurement.
12-12
Components & Accessories