Horiba Scientific FluoroMax-4 Operation Manual page 45

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
i
j
k
l
m
n
Cursor's wave-
length in nm
Enter the current, observed position (here, 477 nm) of the peak in the
Position Control field, then hit the Enter key.
Click the Calibrate Excitation 1 button.
The Calibrate window opens:
In Peak Of
Interest, enter
the actual or
expected
position of the
peak (it ought
to be 467 nm).
Click the OK
button.
At the bottom right of the Real Time Control window, click the Cancel
button.
In the Experiment window, click the Run button to confirm the correct
peak position.
A correct scan is shown below (peak is at 467 nm):
Cursor
3-9
System Operation