Horiba Scientific FluoroMax-4 Operation Manual page 36

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Electronics and controllers (1a)
The rear and bottom of the FluoroMax
instrument, scans, and measurements.
Lamp power supply (1a), optional phosphorimeter power supply, and side-panel
circuit board.
Xenon-
lamp power
supply (1a)
J400906
board
Computer system and software (not on diagram)
Not shown on the schematic is the host computer with FluorEssence™ software. The
technical specifications chapter lists the computer requirements. An optional printer or
network card is useful for printing. FluorEssence™ software for Windows
interaction with the spectrofluorometer. For information on FluorEssence™, see the
FluorEssence™ User's Guide and the on-line help files within FluorEssence™.
®
-4 houses the electronics for running the lamp,
Xe-lamp power supply (1a)
Optional phosphorimeter-
lamp power supply
This supply is a tunable 180-W-maximum power supply. It supplies
a large start-up voltage to the 150-W xenon lamp, then holds the
lamp steady at 15 V. The voltage is filtered, to stabilize the illumina-
tion as much as possible. The lamp is started with the rocker-switch
on the right panel of the instrument, and controlled through software.
Underneath the optical platform, there is a control board. This con-
trols the monochromators and any optional accessories connected to
the sample compartment.
2-6
System Description
Side-panel circuit board
®
controls all