Horiba Scientific FluoroMax-4 Operation Manual page 183

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Weight
Ambient temperature
range
Maximum relative
humidity
Power
Fuses
Phosphorimeter (FluoroMax
The following components and specifications also apply to the FluoroMax
®
FluoroMax
-P with FM-2005 upgrade.
Source
Flash rate
Lifetime range
Flash duration
Delay after flash
Flashes per data
point
Sample window
TCSPC option (FM4-2015 upgrade)
The following components and specifications also apply to a FluoroMax
2015 upgrade.
Source
Lifetime range
Minimum resolution 55 ps/channel
Detection method
75 lbs (34 kg)
15–30°C
59–86°F
75%
Universal AC single-phase input power; 85–250 V AC; line
frequency 50–60 Hz.
Two 5 × 20 mm IEC approved, 4.0 A, 250 V, Time Delay fus-
es (Cooper Bussman part number GDC-4A or equivalent)
®
-4P)
UV xenon flash tube
0.05–25 Hz
Down to 10 μs
3 μs at full-width half-maximum. Low-intensity tail extends >
30 μs.
50 μs to 10 s, in increments of 1 μs.
1–999
10 μs to 10 s, in increments of 1 μs.
Interchangeable NanoLEDs; peak wavelengths 265 nm–785 nm.
200 ps–0.1 ms
Time-correlated single-photon counting (TCSPC)
11-3
Technical Specifications
®
-4P, or
®
-4 with FM4-

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fluoromax-4p

Table of Contents