Hardware Init. Error
appears.
Sample turret is not
operating.
®
FluoroMax
sets itself in
phosphorimeter
mode.
"Data file does not
exist" or "file read
error" message ap-
pears
Broken IR sensor in
monochromator.
Software is not enabled.
USB cable is connected
improperly.
Flash lamp's life is ex-
-4P re-
pired, drawing excess
current, causing transient
emf interference.
User is not logged into
®
Windows
XP as admin-
istrator or power user
Replace IR sensor: Call the Ser-
vice Department.
Check status.
Check USB cable's connection.
Replace xenon flash lamp.
Log into Windows
istrator and or power user, and
restart FluorEssence™
7-3