Horiba Scientific FluoroMax-4 Operation Manual page 112

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
6
Acquire another water Raman scan.
a
b
c
No further adjustment is
necessary.
Replace the lamp-housing cover.
Replace the FluoroMax
STOP.
®
The FluoroMax
-4 is ready for operation.
Use the same parameters as in step 3.
Note the peak intensity.
Use the flowchart below.
Is the signal
Yes
within speci-
fications?
®
-4 cover.
No
Continue to adjust the
screws, to improve the
water Raman signal.
After repeated
No
adjustments, is
the signal still
low?
Go to Chapter 7: Troubleshooting.
6-10
Maintenance
Yes