Horiba Scientific FluoroMax-4 Operation Manual page 157

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
c
3
Gently remove the
sample mount from
the front of the
FluoroMax
The standard FluoroMax
front is held via a friction fit,
with no screws to remove. Some
accessories require removal of 4
screws. Some sample mounts al-
so have a 15-pin connector at
the inside end for automated ac-
cessories.
4
Remove the lamp cover.
a
b
Disconnect the RS-232 cable, power cord, and any other cables attached
to the spectrofluorometer.
®
-4P.
®
-4P
With an Allen key, remove
the five screws from inside
the left wall of the sample
compartment.
Pull the lamp
cover to the left
about 2 cm.
FluoroMax
9-13
®
-4P Phosphorimeter Operation