Horiba Scientific FluoroMax-4 Operation Manual page 26

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
Directions
1
Unpack and set up the FluoroMax
a
b
c
d
e
f
g
h
i
j
2
Set up the computer.
The information gathered by the
spectrofluorometer system is displayed
and controlled through the host PC via
FluorEssence™ software. The host PC
Carefully open the FluoroMax
Remove the foam-injected top piece and any other shipping restraints in
the carton.
With assistance, carefully lift the instrument from the carton, and rest it
on the side of the laboratory bench where the system will stay.
Place the instrument in its permanent location.
Level the spectrofluorometer.
Adjust the four leveling feet on the bottom of the instrument.
Inspect for previously hidden damage.
Notify the carrier and HORIBA Scientific if any is found.
Check the packing list to verify that all components and accessories are
present.
Plug one end of the power cord into the proper receptacle on the right
side (while facing the unit) of the spectrofluorometer.
Plug one end of the USB cable
into the USB receptacle.
With an optional trigger
accessory, plug one end of the
trigger cable into the TRIGGER
connector on the FluoroMax
Allow the unconnected ends of
the cables to dangle freely; they
will be connected in later steps.
®
®
-4 shipping carton.
Caution:
Watch your fingers!
®
-4.
1-6
Requirements & Installation
.