Activate the Subtract Dark checkbox.
Activate the Remeasure Anisotropy Only checkbox.
Activate Reset to Mechanical Zero only if the polarizers are definitely
miscalibrated. This deletes the previous calibration.
5
Place the Ludox
holder.
6
Click the
Continue
button.
The system rotates through the po-
larizers as shown in the checklist on the window. As each phase is completed,
the checkboxes are up-
dated. If the sample is too
concentrated or dilute,
the software prompts you
to correct this.
When complete, the
software displays the measured anisotropy for the emission channel (S).
®
or glycogen in the sample
Caution:
Safety Data Sheets (MSDS)
for hazards regarding the use
of glycogen or colloidal silica.
10-8
Automated Polarizers
Refer to your Material