Horiba Scientific FluoroMax-4 Operation Manual page 141

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jun 2012)
4
Choose Preferences.
The Preferences area appears.
5
Choose the Instrument Correction Files icon.
The Instrument Correction Files area appears.
6
If there are no active fields in the Instrument
Correction Files area, click the Insert button.
7
In sequence,
a
b
c
Choose the
Detector
column, and
select the appropriate detector from the drop-down list,
The Mono column, and the correct monochromator from the drop-down
list,
The Grating column, the correct grating from the drop-down list,
Note:
The signal detector, S, uses only
the mcorrect file; the reference detec-
tor, R, uses only the xcorrect file.
8-11
Producing Correction Factors