Horiba Scientific FluoroMax-4 Operation Manual page 138

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jun 2012)
b
3
Export the normalized
correction-factor
graph.
In the File menu, click Export, and
select SPC Data....
4
Apply the correction
factor into your sys-
tem configuration by
following the instruc-
tions in your instru-
ment manual.
Remove the accessory.
1
Remove the sample-compartment cover.
Save this new file as irrad2.
This normalizes the correction factor file so that the minimum intensity
–1
will be 1 count s
. Highlight the B(Y) column, right-click and select
Normalize.
Warning:
Wear appropriate eye-
protection against UV, visible, and
IR when the tungsten-halogen lamp
is on and the sample-compartment
is uncovered.
8-8
Producing Correction Factors