Horiba Scientific FluoroMax-4 Operation Manual page 73

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jul 2012)
2
Connect to the desired instrument configuration.
a
In the System Configuration area,
select the configuration from the
drop-down menu.
b
Click the Connect Config button
to initialize the instrument.
3
Activate the excitation and emission
monochromators' checkboxes.
4
Activate the detector checkbox(es)
and switch on high voltage if
necessary, in the Detectors
area.
5
If desired, in the Ratios area,
activate the ratio S/R checkbox to obtain cor-
rected output.
6
In the Time area, choose
various wavelength-groups
and integration time.
a
For multiple integration times, use the
Time / Accessories tab to add multiple
time intervals.
b
Choose the wavelength pair or multiple
wavelength sets in the Monos /
Detectors tab.
7
Select the accessories, looping, and delays.