Excitation-monochromator calibration check
This calibration check verifies the wavelength calibration of your excitation
monochromator, using the reference photodiode located before the sample compart-
ment. It is an excitation scan of the xenon lamp's output, and should be the first check
performed.
1
Close the lid of the sample compartment.
2
On the main FluorEssence toolbar, select the
Experiment Menu button
The Fluorescence Main Experiment Menu appears:
3
Choose Spec-
tra.
The Experiment Type window appears:
4
Choose Excitation.
The xenon-lamp scan experiment au-
tomatically loads:
:
3-4