Horiba Scientific FluoroMax-4 Operation Manual page 142

With usb
Table of Contents

Advertisement

®
®
FluoroMax
-4 & FluoroMax
-4P with USB rev. D (30 Jun 2012)
d
8
When all necessary detectors have an associat-
ed correction-factor file, click the OK button.
9
Reinitialize the instrument and software.
a
b
c
The new correction-factor files are now ready to be activated in the Experiment
Setup window, under the Detectors icon:
Then, in the File column, browse for the appropriate correction-factor
file.
In the FluorEssence™ toolbar, choose Collect.
In the drop-down menu, choose Advanced Setup.
In the drop-down menu, choose System ReInitialization.
8-12
Producing Correction Factors